Mechanical Information - electronics module
Page 7 of 10
.75 [19.1]
2.03 [51.5]
2.18 [55.4]
1.89 [48.0]
1.50 [38.1]
Ø.13 [Ø3.3]
2.11 [53.5]
.10 [2.4]
.85 [21.5]
.46 [11.6]
PIN 1
PIN 5
PIN 10
PIN 15
Male D-sub connector
Maximum Quadrature Output Frequency
Output Frequency A-quad-B Output Rate Dwell Time (or edge separation)
7.2MHz 28.8M quadrature states / sec 0.035µsec
3.6MHz 14.4M quadrature states / sec 0.069µsec
1.8MHz 7.2M quadrature states / sec 0.139µsec
900kHz 3.6M quadrature states / sec 0.278µsec
450kHz 1.8M quadrature states / sec 0.555µsec
225kHz 900k quadrature states / sec 1.111µsec
All Specifications are subject to change. All data is accurate to the best of our knowledge.
MicroE Systems is not responsible for errors.
Mercury
3000
encoder
120
Ω
Cable Zo=120Ω
Customer
Electronics
Signal Termination for
A-Quad-B and Index
Output Signals
A+
B+
I+
Pins 4 & 5
Pins 9 & 10
Pins 14 & 15*
Quadrature TTL
*Note: The index pulse may be aligned with A- or B- at some interpolation values.
PIN 11